Sysmex XS-1000iC hematology analyzer 510(k) document confirming MCV stability up to 48 hours.
This document is a 510(k) Substantial Equivalence Determination for the Sysmex XS-1000iC Automated Hematology Analyzer, confirming its extension of Mean Cell Volume (MCV) stability to 48 hours. It details the device’s system components, operation modes, calibration, and performance characteristics including accuracy, precision, linearity, and stability.
The analyzer uses flow cytometry and hydrodynamic focusing methods to analyze 21 blood parameters. It includes an auto loader and information processing unit, with optional graphic printer and barcode reader. The device is classified as Class II and is intended for in vitro diagnostic use in clinical and reference laboratories.
| MCV Stability | 48 hours |
|---|---|
| Parameters Analyzed | 21 parameters |
| Classification | Class II |
| Product Code | GKZ |
| Regulation Section | 21 CFR 864.5220 |
The MCV parameter is stable up to 48 hours for samples stored at room temperature and 4ºC.
It analyzes 21 parameters including WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, MPV, RDW-SD, RDW-CV, NE UT%, NEUT#, LYMPH%, LYMPH#, MONO%, MONO#, EO%, EO#, BASO%, and BASO#.
The device is classified as Class II.
The predicate device is the Sysmex® XE-2100DC, with 510(k) number K051459.