This manual describes how to use the PCE-CT 65 Coating Thickness Gauge, which can measure the coating thickness of ferrous and non-ferrous metals.
The PCE-CT 65 is a handheld coating thickness gauge that uses magnetic induction for ferrous metals and eddy current for non-ferrous metals. It offers a measuring range of 0–1350 µm (0–53.1 mils) with selectable units, data storage in 30 working modes, and USB connectivity for data transfer.
The manual provides safety instructions, technical specifications, device and display descriptions, battery replacement, measurement procedures, settings (units, backlight, auto power-off), calibration steps for zero points on both metal types, troubleshooting, warranty, and disposal information. This manual is written in Spanish.
| Sensor type | F (ferrous) and N (non-ferrous) |
|---|---|
| Measuring principle (F) | Magnetic induction |
| Measuring principle (N) | Eddy current |
| Measuring range | 0 … 1350 µm (0 … 53.1 mils) |
| Accuracy (0–1000 µm) | ±(2.5% + 2 µm) |
| Accuracy (1000–1350 µm) | ±3.5% |
| Resolution (0–100 µm) | 0.1 µm |
| Resolution (100–1000 µm) | 1 µm |
| Resolution (1000–1350 µm) | 0.01 µm |
| Minimum curvature radius (F) | 1.5 mm |
| Minimum surface diameter (F) | 7 mm |
| Smallest measurable thickness (F) | 0.5 mm |
| Operating conditions | 0 … +40 °C, 20 … 90% RH |
The gauge measures coating thickness from 0 to 1350 µm (0 to 53.1 mils) on both ferrous and non-ferrous metals.
Zero calibration for ferrous and non-ferrous metals is described in section 6.1; zero points can be deleted in section 6.2.
The device operates from 0 to +40 °C and 20 to 90% relative humidity.
The gauge has 31 working modes; mode 0 does not store data, while the other 30 modes can each store up to 30 readings in a circular memory.