This document describes the use of Dionex IonPac AS7 and AG7 ion chromatography columns.
The IonPac AS7 analytical column with the AG7 guard column is designed for the analysis of a wide range of polyvalent anions, including polyphosphates, polyphosphonates, cyanide, and iodide. It is compatible with pH 0-14 eluents and eluents containing organic solvents from 0-5% in concentration.
The manual covers system requirements, installation, operation, example applications, troubleshooting, quality assurance reports, column care, and system configuration for 2-mm and 4-mm formats.
| Particle size | 10.0 µm |
|---|---|
| Cross-linking | 2% |
| Ion-exchange capacity | 100 µeq/column (AS7 4-mm) |
| Functional group | Alkyl quaternary ammonium |
| Hydrophobicity | Medium-High |
| Typical back pressure | 900 psi (6.20 MPa) for AS7 4-mm |
| Typical flow rate | 0.5 mL/min for 4-mm; 0.12 mL/min for 2-mm |
| Maximum flow rate | 3.0 mL/min for 4-mm; 0.75 mL/min for 2-mm |
| Eluent pH range | 0-14 |
| Organic solvent content in eluents | 0-5% |
They are designed for a wide variety of applications, including polyphosphates, polyphosphonates, cyanide, and iodide.
Eluents must contain less than 5% organic solvents.
The AS7 is compatible with pH 0-14 eluents.