This product manual provides instructions for using the Dionex MONOLITH ANION CONCENTRATOR (MAC).
The IonSwift Monolith Anion Concentrator (MAC) columns, specifically the MAC-100 and MAC-200, are designed to support ICS-5000 Capillary systems. They function by stripping ions from a measured volume of a relatively clean aqueous sample matrix, concentrating the desired analyte species to lower detection limits by 2-5 orders of magnitude. The unique monolith media allows for direct connection as a loop to the injection valve without end fittings or couplers.
| MAC-100 Column Dimensions | 0.5 x 80mm |
|---|---|
| MAC-200 Column Dimensions | 0.75 x 80mm |
| MAC-100 Maximum Flow Rate | 0.2 mL/min |
| Maximum Operating Pressure | 3,000 psi (20.7 MPa) |
The recommended start-up flow rate for the MAC-100 is less than 0.05 mL/min.
The recommended start-up flow rate for the MAC-200 is less than 0.5 mL/min.
For long-term storage, use 100 mM Sodium Borate prepared from Na2B4O7*10 H2O.