The ion chromatograph manual introduces the composition, control, instructions, safety signs, deionized water requirements, instrument description, configuration, startup, operation, shutdown, maintenance, fault guide, and maintenance of the instrument.
The Dionex ICS-5000 Ion Chromatography System is designed for both capillary and analytical applications, featuring dual pumps, a degasser, and various detector options including conductivity and electrochemical detectors. The system can be controlled using the Chromeleon 7.0 or 6.8 software platform running on Windows XP or Vista.
The manual details the system's components, including the pump types (equal concentration microflow, equal concentration analytical, and four-port gradient analytical), the eluent generator (EG), and the detector/chromatography unit (DC). It also covers the specifications for deionized water requirements, safety warnings, and specific operational steps for manual and automatic injection. This manual is written in Chinese.
| Flow Rate Range (Microflow Pump) | 0.0001 ~ 4.0mL/min |
|---|---|
| Flow Rate Range (Analytical Pump) | 0.001 ~ 10.0mL/min |
| Maximum Operating Pressure | 35MPa (5000psi) |
| Power Supply | 90 ~ 265V, 47 ~ 63Hz |
| Operating Temperature Range | 4 ~ 40°C |
| Relative Humidity | 5 ~ 95% |
The system can be controlled using Chromeleon 7.0 SR1 or Chromeleon 6.8 0 SR9.
The water must have a resistivity greater than 18MΩ, TOC less than 10ppb, and metal ions less than 1ppb.
The SP pump offers equal concentration microflow, equal concentration analytical, and four-port gradient analytical types.
The maximum operating pressure is 35MPa (5000psi).