The Sartorius Pipette Cleaning and Decontamination Guide provides instructions on how to clean and decontaminate Sartorius pipettes.
This Sartorius information guide describes cleaning and decontamination procedures for Sartorius pipettes. It explains the difference between cleaning and decontamination, gives daily cleaning steps for the outer surface, and covers cleaning of the lower part every three months. It also addresses disassembly and reassembly for mechanical pipettes, including Proline, and notes that multichannel lower parts should be serviced by an authorized Sartorius service provider.
The guide includes application-specific cleaning and decontamination reagent guidance, autoclaving sections for mechanical and electronic pipettes, and a summary of pipette parts and materials for Tacta, mLine, Proline Plus, Proline, Picus, and Picus NxT. It also gives contact details for further information and for maintenance, servicing, or calibration.
| Small-volume pipettes | 3, 10, 20, and 100 µl |
|---|---|
| Mid-volume pipettes | 200 and 1000 µl |
| Large-volume pipettes | 5000 µl and 10 ml |
The guide recommends cleaning and decontaminating it and checking performance every three months, and also after over-aspiration or clear contamination.
It lists 70% (v/v) ethyl alcohol for both the upper and lower parts of the listed pipettes.
Only an authorized Sartorius service provider should do this.
The guide recommends calibration at minimum once a year.