Application note describing how to connect BioPAT® MFCS 4.8 and above to an Endress+Hauser Raman Rxn analyzer via OPC UA and acquire process data.
This application note explains how to connect a BioPAT MFCS system (version 4.8 and above) to an Endress+Hauser Raman Rxn analyzer via OPC UA. It covers software, hardware and license prerequisites, specific connection parameters, OPC UA namespace information, and unit configuration in BioPAT MFCS.
The document is intended for customers who want to use Raman Rxn univariate data in BioPAT MFCS for automated process control, process monitoring and alarm limits. It notes that applying a SIMCA model and managing the device are outside its scope.
| BioPAT MFCS software version | v4.8 or above |
|---|---|
| Raman RunTime software version | v6.2.2 or above |
| Connection type | Ethernet based connection |
| Required BioPAT MFCS license | OPC Client Module |
| Raman Rxn authentication mode | Username |
| Default Raman Rxn OPC UA username | kaiser-opc |
| Default Raman Rxn OPC UA password | opc |
BioPAT MFCS v4.8 or above is recommended for connection to a Raman Rxn analyzer.
The BioPAT MFCS OPC Client Module must be licensed.
No. The authentication mode is Username, because the Raman Rxn device does not support anonymous access.
The default OPC UA username is kaiser-opc.