About this manual

The Virosart HF virus filter family utilizes an asymmetric polyether-sulfone (PES) hollow-fiber membrane with a nominal pore size of 20 nm. It is designed for the virus filtration of monoclonal antibodies, antibody fragments, and small recombinant proteins towards the end of the purification process. The membrane features a funnel-like pore-size gradient and is surface modified with a hydrogel-forming, low binding polymer to ensure robust retention of small viruses while maintaining high product transmission.

What's inside

  • Virus Filtration
  • Product Overview
  • Product Release
  • Throughput
  • Filtration Basics
  • Filtration of mAbs
  • Consistent Filter Throughput
  • Protein Transmission
  • Scalability
  • Virus Clearance

Specifications

Membrane TypeAsymmetric polyether-sulfone (PES) hollow-fiber membrane
Nominal Pore Size20 nm
Target Product Size< 150 kD
Water Flow Rate Test Pressure2.0 bar | 30 psi
Integrity Test Pressure4.5 bar | 65.25 psi
Integrity Test Duration5 min
Release Criteria VirusPseudomonas aeruginosa bacteriophage PP7

Frequently asked questions

What is the main application for Virosart HF filters?

The main application is the virus filtration of monoclonal antibodies, antibody fragments, or small recombinant proteins towards the end of the purification process.

How is the Virosart HF membrane surface modified?

The membrane is surface modified with a hydrogel-forming, low binding polymer to reduce the adsorption of proteins and other potentially fouling species.

What is the release criteria for Virosart HF membranes?

The release criteria is the retention of Pseudomonas aeruginosa bacteriophage PP7, which is used as a standard model virus for small viruses.