The ammonia gas-sensitive electrode is a gas-permeable membrane ion-selective electrode used to test free ammonia in aqueous solutions, which can be simple, fast, accurate and economical.
This document is the user manual for the Ohaus STISE37 composite ammonia gas-sensitive electrode. It describes the electrode's specifications, including its temperature range, cable length, and measurement range for free ammonia in aqueous solutions. The guide explains that the electrode includes a reference electrode and is designed for simple, rapid, and precise testing.
The manual details the operational procedures, including the required equipment (such as the ST5000i ion meter), preparation steps like filling the reference liquid and activating the electrode with 2M NH4Cl solution, and performance checks. It also covers sample preparation, measurement protocols using standard solutions, and methods for storing and maintaining the electrode to ensure accuracy and longevity. This manual is written in Chinese.
| Temperature Range | 5~40 °C |
|---|---|
| Reference Electrode Type | Fillable |
| Cable Length | 1 m |
| Electrode Rod Length | 120 mm |
| Electrode Rod Diameter | 12 mm |
| Electrode Resistance | <500 MΩ |
| Measurement Range | 1 M~5 × 10-6 M |
| Minimum Sample Volume | 50ml in beaker |
If calibrated once per hour, the direct measurement repeatability is approximately ±4%.
Measuring high concentration solutions requires approximately 1 minute to reach 99% of the stable potential value.
The difference between the two readings (E2-E1) should be -56mV±4mV at 25°C.