Ohaus STISE37 Ammonia Gas-Sensitive Electrode User Manual

The ammonia gas-sensitive electrode is a gas-permeable membrane ion-selective electrode used to test free ammonia in aqueous solutions, which can be simple, fast, accurate and economical.


Ohaus STISE37 Ammonia Gas-Sensitive Electrode User Manual - cover page
Brand
OHAUS
Category
Other Electronics
Document type
User Manual
Language
Chinese
Pages
11
File format
PDF
File size
655 KB
Published
15 November, 2023
Updated
15 November, 2023
MD5 checksum
0A0DA8AA1E4A92287CBC745630D365DB

About this manual

This document is the user manual for the Ohaus STISE37 composite ammonia gas-sensitive electrode. It describes the electrode's specifications, including its temperature range, cable length, and measurement range for free ammonia in aqueous solutions. The guide explains that the electrode includes a reference electrode and is designed for simple, rapid, and precise testing.

The manual details the operational procedures, including the required equipment (such as the ST5000i ion meter), preparation steps like filling the reference liquid and activating the electrode with 2M NH4Cl solution, and performance checks. It also covers sample preparation, measurement protocols using standard solutions, and methods for storing and maintaining the electrode to ensure accuracy and longevity. This manual is written in Chinese.

What's inside

  • Introduction and Ordering Information
  • Electrode Parameters and Characteristics
  • Required Equipment and Solutions
  • Electrode Preparation
  • Checking Electrode Performance
  • Sample Requirements and Measurement
  • Storage and Maintenance
  • Frequently Asked Questions

Specifications

Temperature Range5~40 °C
Reference Electrode TypeFillable
Cable Length1 m
Electrode Rod Length120 mm
Electrode Rod Diameter12 mm
Electrode Resistance<500 MΩ
Measurement Range1 M~5 × 10-6 M
Minimum Sample Volume50ml in beaker

Frequently asked questions

How often should the electrode be calibrated?

If calibrated once per hour, the direct measurement repeatability is approximately ±4%.

What is the expected response time for high concentration solutions?

Measuring high concentration solutions requires approximately 1 minute to reach 99% of the stable potential value.

What is the expected slope difference for a qualified electrode?

The difference between the two readings (E2-E1) should be -56mV±4mV at 25°C.