The capaNCDT 6200 series sensors are non-contact sensors for measuring distance, position and thickness. They are characterized by their high accuracy, high resolution and fast response time.
This quick manual from MICRO-EPSILON MESSTECHNIK GmbH & Co. KG covers the capaNCDT 6200 series capacitive measuring systems with the controller models DT6220, DT6222, DT6228, DT6230 and DT6238. It describes the intended use of the system for measuring displacement, distance, movement, thickness and the position of parts or machine components in an industrial environment.
The document explains setup and connection options, grounding of the target, synchronization between controllers using the SC6000-x cable, sensor and controller assembly, insertion of the demodulator module, commissioning including IP address configuration, specifying sensor measuring ranges, the math function for scaling and linking channels, distance measurements, and operation and maintenance. It also lists warnings, safety symbols and the manufacturer's disclaimer, and refers readers to the full operating instructions available on the Micro-Epsilon website.
| Controller operating temperature | +10 ... +60 °C (+50 ... +140 °F) |
|---|---|
| Controller storage temperature | -10 ... +75 °C (+14 ... +167 °F) |
| Sensor operating temperature (CS005 to CS10) | -50 ... +200 °C (-58 ... +392 °F) |
| Protection class | IP40 |
| Humidity | 5 ... 95 % (non-condensing) |
| Ambient pressure | Atmospheric pressure |
| Data channel value range | 0 to 1000000 µm |
| Math function factor range | -9.9 to +9.9 |
With two synchronized DL62xx demodulators, the target does not need to be grounded in most cases thanks to Micro-Epsilon's synchronization technology.
Plug the SC6000-x synchronization cable into the SYNC OUT port on Controller 1 and into the SYNC IN port on Controller 2; each controller can be the master.
Go to the Settings > Channel n > Channel information menu and specify the measuring range of the sensor manually.
Turn off the supply voltage before cleaning, clean with a damp cloth and then rub the sensor surface dry.