MICRO-EPSILON thermoMETER CT Infrared Sensor User Manual

User manual for MICRO-EPSILON's infrared temperature sensors, covering safety, installation, operation, and software usage for CT, CTF, CTH, CTP, CTM-1…


MICRO-EPSILON thermoMETER CT Infrared Sensor User Manual - cover page
Brand
MICRO-EPSILON
Document type
User Manual
Language
English
Pages
98
File format
PDF
File size
1.4 MB
Published
30 October, 2026
Updated
30 October, 2026
MD5 checksum
1DDEAE75514038969F655D2F4C236864

About this manual

This manual provides detailed instructions for installing, operating, and maintaining MICRO-EPSILON's thermoMETER CT infrared sensors. It covers safety precautions, technical specifications, mechanical and electrical installation, sensor setup, and software operation including CompactConnect. The document includes model-specific data for CT, CTF, CTH, CTP, CTM-1, CTM-2, and CTM-3 series, along with emissivity tables and smart averaging features for accurate temperature measurement.

The manual also addresses environmental conditions, cable handling, interface connections (USB, RS232, RS485, CAN, Profibus, Ethernet), relay outputs, functional inputs, and troubleshooting. It includes safety warnings for handling, storage, and operation to prevent damage or injury, and provides guidelines for proper use in industrial and laboratory environments.

What's inside

  • Safety
  • Technical Data
  • Delivery
  • Optical Charts
  • CF Lens and Protective Window
  • Mechanical Installation
  • Electrical Installation
  • Outputs and Inputs
  • Operating
  • CompactConnect Software

Specifications

Protection ClassIP 65 (NEMA 4)
Operating Temperature0 ... 85 °C (+32 ... +180 °F)
Storage Temperature-40 ... 85 °C (-40 ... +185 °F)
Relative Humidity10 ... 95 %, non-condensing
Sensor MaterialStainless steel
Controller MaterialDie casting zinc
Cable Length1 m (CT-SF02, CT-SF15, CT-SF22), 3 m (CTH, CTM-1, CTP), 8 m, 15 m
Power Supply8 - 36 VDC
Current Drawmax. 100 mA
OutputsChannel 1: 0/4-20 mA, 0-5/10 V, thermocouple (J or K) or alarm output; Channel 2: 0-5 V or 0-10 V or alarm output
Digital InterfacesUSB, RS232, RS485, CAN, Profibus DP, Ethernet
Relay Outputs2 x 60 VDC/42 VAC RMS, 0.4 A; potential free
Functional InputsF1 up to F3; software programmable for external emissivity adjustment, ambient temperature compensation, trigger
EMC ComplianceEN 61326-1: 2006 and EN 61326-2-3: 2006

Frequently asked questions

What is the operating temperature range for the controller?

0 to 85 °C (+32 to +185 °F).

What interfaces are available for communication?

USB, RS232, RS485, CAN, Profibus DP, and Ethernet (optional plug-in modules).

What is the power supply voltage for the sensor?

8 to 36 VDC.