TANITA Body Composition Analyzers have been clinically proven to be accurate, reliable and provide highly repeatable results. Our Analyzers are used worldwide by health, research and medical professionals primarily in the following fields: - medical screening and health assessments of adults and children - monitoring the progress of weight loss during medical treatment relating to lifestyle diseases such as diabetes, hyperlipidemia, bariatric surgery, hypertension and fatty liver disease. - monitoring increases of muscle mass, reduction of body fat and hydration levels as part of a fitness or training program - assessing the true effectiveness of nutrition and physical activity programs where body mass index cannot identify key changes in body composition - collating subject data
This manual is for the TANITA MC-980MA-N plus CiD body composition analyzer. It describes intended use in medical screening, health assessments, weight-loss monitoring, fitness programs, and research, and lists safety precautions, delivered parts, assembly, connections, power supply, and setup steps.
It also covers database management, password changes, measurement flow, body-composition and weight-only measurements, troubleshooting, regular maintenance, and specifications such as operating conditions and measurement ranges.
| Usage temperature range | 25 to 35 C |
|---|---|
| Usage relative humidity range | 30 to 80% |
| Maximum altitude | 2,000 m ASL |
| Usage atmospheric pressure range | 86 to 106 kPa |
| Storage temperature range | 10 to 50 C |
| Storage relative humidity range | 10 to 90% |
| Storage atmospheric pressure range | 70 to 106 kPa |
| Age range | 5 to 99 years |
| Height range | 90.0 to 249.9 cm |
| Weight range | 2.0 to 300.0 kg |
| Body fat percentage range | 3.0 to 75.0% |
| Visceral fat rating range | 1 to 59 |
| Bioelectrical data | Reactance/Resistance/Phase Angle |
It is used by health, research, and medical professionals for medical screening and health assessments of adults and children, among other applications.
No. The device must not be used on subjects with pacemakers or other mechanical implants because it passes a weak electrical current through the body.
TANITA recommends periodic checks: check that the unit is on a stable, level surface and verify date and time settings at least daily, inspect the display, cables, safety labels, and accessories at least weekly, and inspect mounting screws at least monthly.
Avoid measuring after strenuous exercise, over-eating, over-drinking, or severe dehydration, and use the equipment under the same conditions and position.