SUN Li-Ion G4 RIC Hearing Aids User Manual

The document is a user manual for hearing aids, which introduces the features and how to use them.


SUN Li-Ion G4 RIC Hearing Aids User Manual - cover page
Category
hearing aid
Document type
User Manual
Language
English
Pages
19
File format
PDF
File size
7 MB
Published
10 November, 2023
Updated
10 November, 2023
MD5 checksum
E35B4B440DDFEA6F322AB5AB7893016B

About this manual

This user manual describes a receiver-in-canal (RIC) hearing instrument with a permanently built-in lithium-ion power cell that is charged in an inductive charger. It explains the instrument parts and names, the push button controls, available hearing programs, and optional features such as AutoPhone, 2earPhone, a Tinnitus-Noiser and binaural synchronization. The guide states that some described features may be optional and should be confirmed by a Hearing Care Professional.

Daily use sections cover charging, the low power signal, turning the instruments on and off via charger or push button, standby mode, inserting and removing the ear pieces, volume adjustment and program changes. Further sections address telephone use, audio induction loops, maintenance and care, wireless functionality, accessories, operating, transport and storage conditions, troubleshooting, and service and warranty.

What's inside

  • Welcome
  • Components and names
  • Controls
  • Settings
  • Charging
  • Turning on and off
  • Switching to standby mode
  • Inserting and removing the hearing instruments
  • Changing the hearing program
  • On the phone

Specifications

Instrument typeRIC (Receiver-in-Canal)
Power cellPermanently built-in lithium-ion rechargeable battery
Battery runtime per chargeUp to 16 hours of use
Low power warning intervalRepeated every 15 minutes, about 2 hours remaining
Operating temperature (discharging)0 to 50 °C
Operating temperature (charging)10 to 35 °C
Operating relative humidity5 to 93 %
Storage temperature0 to 25 °C
Transport temperature-20 to 60 °C
Minimum charging interval in storageAt least every 6 months
Intended ageNot for children under 3 years or persons with a developmental age under 3 years

Frequently asked questions

How long do the hearing instruments last on a full charge?

A full charge is intended to cover a full day of use, given as 16 hours. If a fully charged instrument does not last that long, the power cell should be replaced by a Hearing Care Professional.

How do I turn the hearing instruments on and off?

Either take them out of the charger to turn them on and place them in the charger to turn them off, or press and hold the push button until the startup melody plays, and press and hold for several seconds to turn off.

How can I tell which hearing instrument is for the left or right ear?

Colored markers indicate the side: a red marker means the right ear and a blue marker means the left ear.

What happens when I hold a telephone to my ear?

If the AutoPhone or 2earPhone function is activated, the hearing program switches automatically to a telephone program when the telephone receiver is close to the ear; 2earPhone sends the call to both ears.