NRGT 26-1 Level Transmitter Instruction Manual. The product is used to measure the level of liquids or solids.
The NRGT 26-1 and NRGT 26-1S are level transmitters for permanent water-level measurement in steam boiler and hot-water systems, as well as in condensate or feedwater tanks. They are compact units combining a level measuring electrode and an electronic insert in the connection housing; they work according to the capacitive measuring process and convert level changes into a 4–20 mA current signal, with the measuring range determined by the electrode rod length.
The manual covers intended use, safety instructions, technical data, nameplate and marking, mounting, electrical connection, basic settings, commissioning, fault indication and remedy, maintenance, and dismantling and disposal. The NRGT 26-1S is approved for marine applications. This manual is written in Spanish.
| Operating pressure | PN 40, 32 bar at 238 °C |
|---|---|
| Mechanical connection NRGT 26-1 | Thread G ¾ A, ISO 228 |
| Mechanical connection NRGT 26-1S | Flange DN 50, PN 40, DIN EN 1092-01 |
| Weight NRGT 26-1 | approx. 1.8 kg |
| Weight NRGT 26-1S | approx. 8.0 kg |
| Power supply | 24 V AC/DC |
| Output signal | 4–20 mA proportional to fill level |
| Maximum load | 500 Ω |
| Protection degree | IP 65 according to DIN EN 60529 |
| Maximum ambient temperature | 70 °C |
It is used for permanent water-level measurement in steam boiler and hot-water systems and in condensate or feedwater tanks.
It provides a 4–20 mA current signal proportional to the fill level.
No. According to EU Directive 2014/34/EU, the equipment must not be used in areas with a risk of explosions.
The maximum permissible ambient temperature is 70 °C.