TSI AEROTRAK 8220 Handheld Optical Particle Counter Operation and Service Manual

8220 AEROTRAK TM Handheld Optical Particle Counter Operation and Service Manual


TSI AEROTRAK 8220 Handheld Optical Particle Counter Operation and Service Manual - cover page
Brand
TSI
Category
Analyzers
Document type
User Manual
Language
English
Pages
55
File format
PDF
File size
1.1 MB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
72AE3FDAA6460C8EDA94F5FCD8C17BF6

About this manual

This Operation and Service Manual, Revision B (January 2007), covers the TSI AEROTRAK Model 8220 Handheld Optical Particle Counter. It provides information on unpacking and parts identification, setup, operation, communication with TRAKPRO software, maintenance, troubleshooting, and specifications.

Safety information includes laser safety, labels, and caution/warning symbols. The manual also describes optional accessories such as the thermal printer and temperature/relative humidity probe.

What's inside

  • Safety Information
  • Unpacking and Parts Identification
  • Set Up
  • Operation
  • Screen Layout and Functionality
  • Communication
  • Maintenance
  • Troubleshooting
  • Specifications

Specifications

Particle Size Range0.3 to 10.0 µm
Number of Bins6
Maximum Concentration2,000,000 #/ft3 (5% coincidence loss)
Pump Flow Rate0.1 cfm (2.83 lpm) ±5%
Laser660 nm 50 mW laser diode
Display3.7 in. (9.4 cm) color LCD with touch screen, 320 x 240
Memory128 MB (over 100,000 sample sets)
Battery Life8 hrs, typical use @ 70°F (21°C)
Dimensions10 in. x 4.5 in. x 3 in.
Instrument Weight with Battery2.2 lbs (1.0 kg)
Power100 to 240 VAC, 50 to 60 Hz
Operating Temperature41 to 95 °F (5 to 35 °C)
Operating Relative Humidity10 to 90 % RH, noncondensing
WarrantyThree years on workmanship and material

Frequently asked questions

What is the particle size range of the Model 8220?

The particle size range is 0.3 to 10.0 µm.

What is the maximum concentration for the Model 8220?

The maximum concentration is 2,000,000 #/ft3 at 5% coincidence loss.

What software is used with the instrument?

The specifications list TRAKPRO Data Analysis Software for the instrument.

How long does the battery last?

Battery life is 8 hours, typical use at 70°F (21°C).