User guide for TSI Model 4046/4146 Primary Calibrators covering unpacking, setup, operation, maintenance, and troubleshooting instructions.
This document serves as the User's Guide for the TSI Model 4046 and Model 4146 Primary Calibrators. It provides comprehensive instructions for unpacking the instrument, identifying parts, and assembling the device with the battery pack. The guide details how to supply power via the battery pack or an optional AC adapter and explains the procedure for connecting filters and flow tubes to the inlet and outlet.
| Measurement Range (Model 4046) | 2.5 to 300 L/min |
|---|---|
| Measurement Range (Model 4146) | 0.01 to 20 L/min |
| Accuracy (Model 4046) | ±2% of reading or 0.05 std. L/min |
| Accuracy (Model 4146) | ±2% of reading or 0.005 std. L/min |
| Operation Temperature Range | 0 to 50°C |
| Storage Temperature Range | -20 to 60°C |
| Warm-up Time | 1 min |
| Power Requirement | 7.5 VDC ± 1.5 V, 300 mA maximum |
No, the calibrator should not be used in the presence of flammable or explosive gases or mixtures because it employs a heated platinum sensor.
No, TSI calibrators are not medical devices under FDA 510k and should not be utilized for human respiration measurements.