Ohaus SD Series Scales Service Manual

The service manual for the SD series scales.


Ohaus SD Series Scales Service Manual - cover page
Brand
OHAUS
Category
Other
Document type
Service Manual
Language
English
Pages
34
File format
PDF
File size
870 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
D54065BCFEA34319517A067FCCC579EB

About this manual

This is the Ohaus service manual for the SD Series scales. It is written for service technicians who perform routine maintenance and service on the scale, and it states that the information needed is contained in five chapters: Getting Started, Troubleshooting, Maintenance Procedures, Testing, and Drawings and Parts Lists. The manual covers the SD35, SD75/SD75L and SD200/SD200L models.

Chapter 1 describes service facility requirements, the tools and test equipment required, the specification table, power supply and controls, basic weighing functions, Dynamic Weighing and calibration. Chapter 2 gives a visual inspection and diagnostic guide plus load cell output voltage and resistance tests and PCB testing. Chapter 3 covers preventive maintenance and replacement of the main PCB, battery and function label. Chapter 4 describes power, overload/underload, calibration retention, repeatability, off-center load and linearity tests, and Chapter 5 contains exploded views and parts lists for the small and large models.

What's inside

  • Introduction and Service Facilities
  • Tools and Test Equipment Required
  • Specifications
  • Operation and Power Supply
  • Basic Functions
  • Dynamic Weighing
  • Calibration
  • Troubleshooting and Diagnostic Guide
  • Checking the Load Cell's Signal
  • Preventive Maintenance and Replacement of Major Components

Specifications

SD35 Capacity x Readability35 kg x 0.02 / 77 lb x 0.05
SD75 / SD75L Capacity x Readability75 kg x 0.05 / 165 lb x 0.1
SD200 / SD200L Capacity x Readability200 kg x 0.1 / 440 lb x 0.2
Weighing Unitskg, lb
Application ModesWeigh, Dynamic Weighing
Calibration Weights (fixed)25 kg, 50 lb / 50 kg, 100 lb / 100 kg, 200 lb
Operating Temperature Range5 °C/41 °F to 40 °C/104 °F at 10 to 80% relative humidity, non-condensing
Stabilization Time3 seconds max
Static Overload Limit (Load Cell)150% capacity
PowerAC adapter 9 V DC / 100 mA or 3 "C" alkaline batteries (not included)
Battery Operating Time360 hours
Auto-Shutoff4 minutes with no activity (battery operation only)

Frequently asked questions

What capacity and readability does the SD35 scale have?

The SD35 has a capacity of 35 kg x 0.02 and 77 lb x 0.05.

How long does the scale run on batteries?

Battery operating time is 360 hours using three "C" alkaline batteries.

How is the calibration function locked?

With the scale powered off, remove the back housing and slide the switch on the PCB marked SW3 to the CAL LOCK position, then reassemble the housing.

What tools and test equipment are required for service?

Standard hand tools, a digital voltmeter (DVM) and an optional load cell simulator.

Download PDF (870 KB)

More OHAUS Manuals