The service manual for the SD series scales.
This is the Ohaus service manual for the SD Series scales. It is written for service technicians who perform routine maintenance and service on the scale, and it states that the information needed is contained in five chapters: Getting Started, Troubleshooting, Maintenance Procedures, Testing, and Drawings and Parts Lists. The manual covers the SD35, SD75/SD75L and SD200/SD200L models.
Chapter 1 describes service facility requirements, the tools and test equipment required, the specification table, power supply and controls, basic weighing functions, Dynamic Weighing and calibration. Chapter 2 gives a visual inspection and diagnostic guide plus load cell output voltage and resistance tests and PCB testing. Chapter 3 covers preventive maintenance and replacement of the main PCB, battery and function label. Chapter 4 describes power, overload/underload, calibration retention, repeatability, off-center load and linearity tests, and Chapter 5 contains exploded views and parts lists for the small and large models.
| SD35 Capacity x Readability | 35 kg x 0.02 / 77 lb x 0.05 |
|---|---|
| SD75 / SD75L Capacity x Readability | 75 kg x 0.05 / 165 lb x 0.1 |
| SD200 / SD200L Capacity x Readability | 200 kg x 0.1 / 440 lb x 0.2 |
| Weighing Units | kg, lb |
| Application Modes | Weigh, Dynamic Weighing |
| Calibration Weights (fixed) | 25 kg, 50 lb / 50 kg, 100 lb / 100 kg, 200 lb |
| Operating Temperature Range | 5 °C/41 °F to 40 °C/104 °F at 10 to 80% relative humidity, non-condensing |
| Stabilization Time | 3 seconds max |
| Static Overload Limit (Load Cell) | 150% capacity |
| Power | AC adapter 9 V DC / 100 mA or 3 "C" alkaline batteries (not included) |
| Battery Operating Time | 360 hours |
| Auto-Shutoff | 4 minutes with no activity (battery operation only) |
The SD35 has a capacity of 35 kg x 0.02 and 77 lb x 0.05.
Battery operating time is 360 hours using three "C" alkaline batteries.
With the scale powered off, remove the back housing and slide the switch on the PCB marked SW3 to the CAL LOCK position, then reassemble the housing.
Standard hand tools, a digital voltmeter (DVM) and an optional load cell simulator.