This product manual describes the Dionx IonPac™ Cation Polisher 061777-01.
This product manual covers the Dionex IonPac CP1 Na+ Form (6 x 16 mm, P/N 064930) and CP2 H+ Form (9 x 24 mm, P/N 064931) Cation Polisher columns. The columns are designed to remove metallic contaminants and multivalent cations such as calcium and magnesium from the sample stream during anion analysis.
The manual describes hardware configuration, installation and operation, recommended operating pressures, regeneration and storage, and troubleshooting. It includes procedures for sample pretreatment with the CP1 and CP2 columns and notes that the product is now sold under the Thermo Scientific brand.
| Column capacity, CP1 Na+ Form | 170 µeq/column |
|---|---|
| Column capacity, CP2 H+ Form | 700 µeq/column |
| Maximum operating pressure | 3,000 psi (20.68 MPa) |
| Maximum recommended flow rate | 5 mL/min |
| CP1 back pressure at 2 mL/min | <10 psi |
| CP2 back pressure at 2 mL/min | <40 psi |
They are designed to remove metallic contaminants and other multivalent cations, such as calcium and magnesium, from the sample stream when pursuing anion analysis.
The CP2 H+ Form Cation Polisher column is recommended for large volume sample preconcentration using an external pump.
The maximum recommended flow rate for both columns is 5 mL/min.
A bi-weekly regeneration protocol is recommended when the number of samples to be analyzed is high.