Dionex IonPac CP1/CP2 Cation Polisher Column 061777-01 Product Manual

This product manual describes the Dionx IonPac™ Cation Polisher 061777-01.


Dionex IonPac CP1/CP2 Cation Polisher Column 061777-01 Product Manual - cover page
Brand
Dionex
Category
Other
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
176 KB
Published
07 November, 2023
Updated
07 November, 2023
MD5 checksum
F7E3DA37241AF2F46277DDD94B9C6E76

About this manual

This product manual covers the Dionex IonPac CP1 Na+ Form (6 x 16 mm, P/N 064930) and CP2 H+ Form (9 x 24 mm, P/N 064931) Cation Polisher columns. The columns are designed to remove metallic contaminants and multivalent cations such as calcium and magnesium from the sample stream during anion analysis.

The manual describes hardware configuration, installation and operation, recommended operating pressures, regeneration and storage, and troubleshooting. It includes procedures for sample pretreatment with the CP1 and CP2 columns and notes that the product is now sold under the Thermo Scientific brand.

What's inside

  • Introduction
  • Installation and Operation
  • Hardware Configuration
  • Sample Pretreatment with CP1 Na+ Form Cation Polisher Column
  • Sample Pretreatment with CP2 H+ Form Cation Polisher Column
  • Recommended Operating Pressures
  • Regeneration and Storage
  • Troubleshooting Guide
  • High Back Pressure
  • Poor Peak Shapes or Recovery

Specifications

Column capacity, CP1 Na+ Form170 µeq/column
Column capacity, CP2 H+ Form700 µeq/column
Maximum operating pressure3,000 psi (20.68 MPa)
Maximum recommended flow rate5 mL/min
CP1 back pressure at 2 mL/min<10 psi
CP2 back pressure at 2 mL/min<40 psi

Frequently asked questions

What are the IonPac CP1 and CP2 Cation Polisher columns used for?

They are designed to remove metallic contaminants and other multivalent cations, such as calcium and magnesium, from the sample stream when pursuing anion analysis.

Which column is recommended for large-volume sample preconcentration?

The CP2 H+ Form Cation Polisher column is recommended for large volume sample preconcentration using an external pump.

What is the maximum recommended flow rate for the CP1 and CP2 columns?

The maximum recommended flow rate for both columns is 5 mL/min.

How often should the CP1 Na+ Form Cation Polisher column be regenerated when sample numbers are high?

A bi-weekly regeneration protocol is recommended when the number of samples to be analyzed is high.

Download PDF (176 KB)

More Dionex Manuals


View all Dionex manuals →