Varian Inova-300 NMR Spectrometer User Notice and Operating Procedures

This is a letter to Varian Inova-300 NMR spectrometer users. The letter informs users that the user training procedures have been substantially changed to make it easier for new users to run the spectrometer.


Varian Inova-300 NMR Spectrometer User Notice and Operating Procedures - cover page
Brand
Varian
Category
Analyzer
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
81 KB
Published
04 November, 2023
Updated
04 November, 2023
MD5 checksum
AFF98CB5D30A09E695CEA97D6CE21FC9

About this manual

This two-page document is a user notice addressed to users of the Varian Inova-300 NMR spectrometers, together with a summary of spectrometer operating procedures. The first page explains that the user training procedures for the Inova-300 spectrometers changed in the summer of 2006, describes the emphasis on minimizing manual manipulation during sample changing, locking and shimming, and lists the requests the facility makes of all users, including those trained before the change.

The second page lists the operating procedures: sample setup with a minimum sample volume of 0.7 mL, experiment setup, instrument operation with the instruction never to adjust the lock phase during routine operation, the state in which the spectrometer should be left after experiments, and troubleshooting steps for locking or shimming problems.

What's inside

  • Changed user training procedures
  • Requests to all users
  • Sample setup
  • Experiment setup
  • Instrument operation and lock phase
  • State to leave the spectrometer in
  • Troubleshooting locking or shimming

Specifications

Minimum sample volume0.7 mL
Sample spinning rate for the standard sampleca. 20 Hz
Probe temperature before leaving the spectrometerwithin +/- 10 ºC of room temperature

Frequently asked questions

What sample volume does the Varian Inova-300 require?

The spectrometers require a minimum sample volume of 0.7 mL; smaller volumes are difficult to lock and shim and are not recommended.

Should the lock phase be adjusted during routine operation?

No. The user should never adjust the lock phase during routine spectrometer operation; if it seems misadjusted, contact PINMRF staff.

How should the spectrometer be left after finishing experiments?

With the standard CDCl3 sample in the magnet at the correct depth, sample spinning ON, the temperature controller OFF, lock ON with correct lock power and gain, and shims checked.

What should I do if locking or shimming is a problem?

Remove your sample, insert the CDCl3 standard sample at the correct depth, make sure spinning is on at ca. 20 Hz, enter the command standard, and check that the sample locks.