YMC CHIRAL NEA series are used for optical isomers. You can choose different columns to meet the needs of most compounds.
This manual covers YMC CHIRAL NEA, CD BR and PREP CD series chiral HPLC columns, including functional groups, pH and temperature ranges, recommended upper temperature, connection type and storage solvent.
It also describes reverse-phase and normal-phase eluents, salt concentration limits, solvent conversion, routine cleaning, pressure limits and sample filtration recommendations. This manual is written in Chinese.
| NEA series functional group | R-(+)1-naphthylethylamine / S-(-)1-naphthylethylamine |
|---|---|
| CD BR series functional group | alpha/beta/gamma-bromocyclodextrin |
| PREP CD series functional group | beta-cyclodextrin |
| Analytical column pressure limit | 20 MPa |
| 10 mm ID column pressure limit | 10 MPa |
| NEA salt concentration limit | below 1.0 M |
| CD BR salt concentration limit | below 0.2 M |
| PREP CD salt concentration limit | below 0.5 M |
| Connection type | WT = Waters type |
The manual recommends 20 MPa for analytical columns.
Use below 1.0 M for NEA, below 0.2 M for CD BR and below 0.5 M for PREP CD.
First replace with a water/organic mixture of the same ratio without salts, then clean with increased organic phase concentration.
Use a miscible solvent such as THF or IPA for at least 10 column volumes.