Triart is a new generation of reversed phase column using newly developed hybrid silica as a matrix, which can be used for most analysis conditions. It has excellent separation performance and higher durability, and is very suitable as the preferred column for the separation of various compounds.
This manual describes YMC-Triart and Accura Triart reverse-phase HPLC and UHPLC columns, including their specifications and recommended operating conditions.
It also provides precautions for column connection and system setup, the storage solvent supplied at shipment, general use precautions, and general column cleaning methods. This manual is written in Chinese.
| Triart C18 particle size | 1.9, 3, 5 µm |
|---|---|
| Triart C18 pore size | 12 nm |
| Triart C18 pH range | 1–12 |
| Common operating temperature | 20–40 °C |
| Triart C18 upper temperature | 90 °C (pH 1–7), 50 °C (pH 7–12) |
Liquid should be passed in the direction indicated by the arrow on the column label.
Filter mobile phase and sample solutions through a membrane of 0.2 µm or smaller.
Use at least 15% methanol; if using a less polar organic solvent than methanol, keep it at 10% or more.