This manual introduces the use, cleaning and preservation methods of YMC-Pack SIL series chromatography columns
This manual provides instructions for the YMC YMC-Pack SIL-06 normal-phase high-performance liquid chromatography (HPLC) column. It details the column's connection types, specifically noting that the 'WT' suffix indicates a Waters-compatible connection. The document outlines the recommended elution solvents, including alkanes, alcohols, and esters, and advises on the use of additives like acetic acid or TFA for ionic compounds to improve peak shape. This manual is written in Chinese.
| Operating pH Range | 2.0 to 7.5 |
|---|---|
| Maximum Temperature | 50°C |
| Recommended Temperature Range | 20°C ~ 40°C |
| Pressure Limit (Analysis Column, L≤150mm) | 20 MPa |
| Pressure Limit (Analysis Column, L=250mm) | 25 MPa |
| Pressure Limit (Semi-prep/Prep Column, ID≥10mm) | 10 MPa |
The 'WT' suffix indicates a Waters-compatible connection type.
Common organic solvents include alkanes (n-hexane, n-heptane), alcohols (methanol, ethanol, isopropanol), ethyl acetate, dichloromethane, and chloroform.
The maximum operating temperature is 50°C.