Elma CA600 600A AC/DC Current Clamp User Manual

The manual introduces the use of Elma CA600 600A AC/DC Current Clamp, including how to use it with a multimeter.


Elma CA600 600A AC/DC Current Clamp User Manual - cover page
Brand
ELMA
Category
multimeter
Document type
User Manual
Language
DA
Pages
4
File format
PDF
File size
368 KB
Published
02 November, 2023
Updated
02 November, 2023
MD5 checksum
0992783FA37DA7210A9B2FCD724B584D

About this manual

The Elma CA600 is a robust AC/DC current clamp for measurements up to 600 A. It is intended for use with a multimeter or another instrument with a millivolt input and provides 1 mV per 1 A.

The manual gives the measurement procedure, notes on DC polarity and zero adjustment, battery replacement instructions, and technical data. It includes Danish/Norwegian and Swedish instructions.

What's inside

  • Foreword
  • Measurement Procedure
  • Notes
  • Battery Replacement
  • Technical Data

Specifications

Measurement range1 - 600 A AC/DC
Frequency range0 - 400 Hz
Output1 - 600 mV (1 mV per 1 A)
DC accuracy+/- 2% + 2 mV
AC accuracy+/- 3% + 2 mV (40-400 Hz)
Jaw opening30 mm
Battery1 x 9 V
Operating temperature0 – 50 °C, 70% RH
Dimensions178 x 70 x 33 mm
Weight290 g

Frequently asked questions

What is the measurement range of the Elma CA600?

It measures from 1 to 600 A AC/DC.

What output does the CA600 give to a multimeter?

It gives 1 mV per 1 A.

What type of battery does the CA600 use?

It uses a 9 V alkaline battery.

What can help if the clamp cannot be zero-adjusted on DC?

Opening the clamp and trying again can help.

Download PDF (368 KB)

More ELMA Manuals


View all ELMA manuals →