This is the instruction manual for the Oral Fluid COVID-19 Antigen Rapid Test Kit.
The ICOV-802H COVID-19 Antigen Rapid Test (Oral Fluid) is a single-use test kit intended to detect SARS-CoV-2 nucleocapsid protein antigens in human oral fluid. The package insert says it is for use by laypersons as a self-test for home and workplace settings, and for symptomatic individuals within the first 7 days of symptom onset.
This instruction manual covers intended use, test principle, reagents, warnings, storage, items provided and not provided, limitations, performance characteristics, and a Q&A section. It states that a positive result is preliminary and should be confirmed by a laboratory PCR test.
| Storage temperature | 35.6-86°F (2-30°C) |
|---|---|
| Kit sizes | 1T, 5T, 10T, 20T kits |
| Sensitivity | 90.1% |
| Specificity | 99.3% |
| Overall accuracy | 97.0% |
| Limit of detection | 8X102 TCID50/ml |
Read results after 15 minutes as long as a colored line has appeared next to the Control region (C); do not read the result after 20 minutes.
A positive result means SARS-CoV-2 antigens are present, so it is very likely you have COVID-19 and the result should be confirmed. Self-isolate according to local guidelines and contact coronavirus testing services for a laboratory PCR test.
It means you are negative or the viral load is too low for the test to recognize; a false negative is possible. If symptoms continue, contact the nearest COVID test center or repeat the test with a new kit after 1-2 days.
Store at 35.6-86°F (2-30°C), do not open the pouch until ready for use, and do not freeze.