COSEL GMA300F Medical AC/DC Power Supply 300W Product Leaflet

GMA300F is a DC power supply in the Toshiba medical electrical equipment series with a rated output power of 300W, suitable for medical devices, meets the IEC60601-1-2 fourth edition standard, has UL62368-1, ANSI/AAMI ES60601-1 C-UL (CSA62368-1, CAN/CSA60601-1), EN62368-1, EN60601-1 third edition and other safety certifications, and provides a five-year warranty.


COSEL GMA300F Medical AC/DC Power Supply 300W Product Leaflet - cover page
Brand
Cosel
Category
AC-DC Power
Document type
Product Leaflet
Language
Chinese
Pages
5
File format
PDF
File size
839 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
4EE9393CD0C9942E4422BE96A603175C

About this manual

COSEL GMA300F is a medical AC/DC power supply with 12 V, 24 V, 48 V, and 56 V output models and a maximum output power of 300 W or 302.4 W depending on model. It is suitable for medical equipment under ANSI/AAMI ES60601-1, EN60601-1 third edition, and IEC60601-1-2 fourth edition, including BF-type medical equipment.

The supply has a 2 x 4 inch standard size, forced-air cooling with an external fan, and optional remote on/off, AUX1 (12 V 1 A), and AUX2 (5 V 1 A). Safety certifications include UL62368-1, C-UL, EN62368-1, and EN60601-1 third edition; conducted noise complies with FCC Part 15 Class B, VCCI-B, CISPR32-B, EN55011-B, and EN55032-B. This manual is written in Chinese.

What's inside

  • Features
  • Specifications
  • Model code
  • Outline drawing
  • Pin assignment
  • Installation method
  • Cooling method
  • Derating curves
  • Basic characteristics data
  • Usage precautions

Specifications

Maximum output power300 W (12/24 V models); 302.4 W (48/56 V models)
Input voltageAC 85-264 V
Input frequency50/60 Hz (45-66 Hz)
EfficiencyUp to 93% typ
Power factor0.95 typ at 115 V; 0.90 typ at 230 V
Leakage current0.13/0.30 mA max
Output voltages12, 24, 48, 56 V
Overcurrent protectionActivates above 105% of rated current
Dimensions50.8 x 37 x 101.6 mm
Weight230 g max
Operating temperature-20 to +70 C
Storage temperature-30 to +75 C
CoolingForced air (external fan required)
Safety certificationsUL62368-1, ANSI/AAMI ES60601-1, C-UL, EN62368-1, EN60601-1 third edition, IEC60601-1-2 fourth edition
Conducted noiseFCC Part 15 Class B, VCCI-B, CISPR32-B, EN55011-B, EN55032-B
Insulation (input to output)AC 4000 V for 1 minute

Frequently asked questions

What medical equipment suitability does the GMA300F have?

It is suitable for medical equipment under ANSI/AAMI ES60601-1, EN60601-1 third edition, and IEC60601-1-2 fourth edition, and for BF-type medical equipment.

Can the GMA300F be operated in parallel?

No, it cannot be operated in parallel.

What cooling is required for the GMA300F?

Forced air cooling with an external fan is required.

What warranty is provided for the GMA300F?

A five-year warranty is provided; see the instruction manual.