This document introduces the features and characteristics of IONPAC ATC-3 2-mm ANION TRAP COLUMN (4 x 35 mm, P/N 059661), IONPAC ATC-3 4-mm ANION TRAP COLUMN (9 x 24 mm, P/N 059660), IONPAC ATC-HC ANION TRAP COLUMN (9 x 75 mm, P/N 059604).
This document is a quickstart manual for the Dionex IonPac ATC-3 (2-mm and 4-mm) and ATC-HC anion trap columns. These columns contain high capacity microporous anion-exchange resin designed to strip trace anionic contaminants from eluents and deionized water, preventing interference in trace anion determinations. The ATC-3 is used with standard 2-mm and 4-mm anion-exchange columns, while the ATC-HC is specifically designed for IC systems equipped with the EG40 Eluent Generator.
| Maximum Operating Pressure | 3,000 psi |
|---|---|
| Column Diameter (ATC-3 2-mm) | 2 mm |
| Column Diameter (ATC-3 4-mm) | 4 mm |
| Column Diameter (ATC-HC) | 9x75 mm |
The maximum recommended operating pressure is 3,000 psi. Operating above this limit can cause irreversible loss of column performance.
The ATC-HC Anion Trap Column is specifically designed to be used with IC systems equipped with the EG40 Eluent Generator.
The recommended sodium hydroxide solution should have iron <5 ppm, chloride < 0.005%, and sodium carbonate < 0.1%.