UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual

This is a user manual for UV-2800/2802/2802S spectrophotometer. It mainly introduces the principle, use, features, technical indicators, accessories, appearance, installation, basic operations and photometer modes of the product.


UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual - cover page
Brand
Unico
Category
Analyzers
Document type
User Manual
Language
Chinese
Pages
60
File format
PDF
File size
2 MB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
F73DBC40B14380E9642BC5172EEB9569

About this manual

This is the Chinese user manual for the UNICO UV-2800 series UV-visible spectrophotometers, covering the UV-2800, UV-2802 and UV-2802S models. It explains the measuring principle (transmittance and absorbance relative to a reference, following the Lambert-Beer law), the intended laboratory applications and the main features of the instrument, followed by the technical specifications, the packing list and the external views of each model.

The rest of the manual is practical: installation and warm-up, the LCD display and keypad, photometer modes (absorbance, transmittance, concentration), quantitative measurement with calibration curves, spectrum scanning, kinetics, DNA/protein measurement, multi-wavelength measurement, system settings and instrument calibration, and special experiments. Appendices cover the DNA/protein test algorithm, lamp replacement and correction methods. This manual is written in Chinese.

What's inside

  • Overview: principle, uses and features
  • Main technical specifications, accessories and appearance
  • Instrument installation
  • Display and keys, power-on and basic operation
  • Photometer mode: absorbance, transmittance, concentration
  • Quantitative measurement and standard curves
  • Spectrum scan
  • Kinetics measurement
  • DNA/Protein measurement
  • Multi-wavelength measurement

Specifications

Spectral bandwidth4 nm (UV-2800), 1.8 nm (UV-2802), 0.5/1/2/4 nm (UV-2802S)
Wavelength range190–1100 nm
Wavelength accuracy±0.8 nm / ±0.5 nm
Wavelength repeatability0.5 nm / 0.3 nm
Stray light≤0.15%T at 220 nm and 340 nm
Photometric range0–200%T, -0.3–2.80A, 0–9999C (0–9999F)
Display320×240 graphic LCD
Data outputRS-232C serial port and Centronics parallel port for HP/Epson compatible laser and inkjet printers
Dimensions550×400×270 mm (UV-2800); 580×400×280 mm (UV-2802/2802S)
Weight16 kg; approx. 21 kg

Frequently asked questions

What is the wavelength range of the UV-2800 series spectrophotometer?

The manual lists the wavelength range as 190 to 1100 nm for the UV-2800, UV-2802 and UV-2802S.

How should the lamp be handled when it is replaced?

Do not handle the lamp with bare fingers; use tissue or cloth, because oil from fingers can make the lamp burn out prematurely.

What output ports does the instrument have for printing or connecting to a PC?

It provides an RS-232C serial port and a Centronics parallel port for HP or Epson compatible laser and inkjet printers.

Download PDF (2 MB)