Key facts
Start here

UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual
This is the Chinese user manual for the UNICO UV-2800 series UV-visible spectrophotometers, covering the UV-2800, UV-2802 and UV-2802S models. It explains the measuring principle (transmittance and absorbance relative to a reference, following the Lambert-Beer law), the intended laboratory applications and the main features of the instrument, followed by the technical specifications, the packing list and the external views of each model.
The rest of the manual is practical: installation and warm-up, the LCD display and keypad, photometer modes (absorbance, transmittance, concentration), quantitative measurement with calibration curves, spectrum scanning, kinetics, DNA/protein measurement, multi-wavelength measurement, system settings and instrument calibration, and special experiments. Appendices cover the DNA/protein test algorithm, lamp replacement and correction methods. This manual is written in Chinese.
UNICO UV-2802 specifications
| Spectral bandwidth | 4 nm (UV-2800), 1.8 nm (UV-2802), 0.5/1/2/4 nm (UV-2802S) |
|---|---|
| Wavelength range | 190–1100 nm |
| Wavelength accuracy | ±0.8 nm / ±0.5 nm |
| Wavelength repeatability | 0.5 nm / 0.3 nm |
| Stray light | ≤0.15%T at 220 nm and 340 nm |
| Photometric range | 0–200%T, -0.3–2.80A, 0–9999C (0–9999F) |
| Display | 320×240 graphic LCD |
| Data output | RS-232C serial port and Centronics parallel port for HP/Epson compatible laser and inkjet printers |
| Dimensions | 550×400×270 mm (UV-2800); 580×400×280 mm (UV-2802/2802S) |
| Weight | 16 kg; approx. 21 kg |
Source: UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual
All UNICO UV-2802 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the Unico UV-2802 used for?
The instrument performs measurements in Basic mode, Quantitative analysis, WL scan, Kinetics, and DNA/Protein analysis.
What are the available measurement modes?
The device offers Basic mode, Quantitative analysis, WL scan, Kinetics, and DNA/Protein analysis.
What is the wavelength range of the UV-2800 series spectrophotometer?
The manual lists the wavelength range as 190 to 1100 nm for the UV-2800, UV-2802 and UV-2802S.
From: UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual
How should the lamp be handled when it is replaced?
Do not handle the lamp with bare fingers; use tissue or cloth, because oil from fingers can make the lamp burn out prematurely.
From: UNICO UV-2800/2802/2802S UV-Visible Spectrophotometer User Manual