This manual provides operating instructions and technical specifications for the UNICO WFJ2100 and WFZUV2100 single-beam spectrophotometers.
This document is the user manual for the UNICO WFJ2100 Visible Spectrophotometer and the WFZUV2100 UV-Visible Spectrophotometer. It explains the operating principles based on the relative measurement method and the application of the Lambert-Beer law. The instruments are designed for scientific research and quality control in various industries, featuring microprocessor technology for automatic settings and data processing. This manual is written in Chinese.
| Optical System | Single beam, 1200 lines/mm diffraction grating |
|---|---|
| Spectral Bandwidth | 5nm |
| Wavelength Range (WFJ2100) | 325—1000 nm |
| Wavelength Range (WFZUV2100) | 200—1000 nm |
| Wavelength Accuracy | ±2.0 nm |
| Stray Light | ≤0.3%T at 360nm (WFJ2100) |
| Photometric Range | 0-125%T, -0.097-2.500A |
| Dimensions | 470mm x 400 mm x 140 mm |
| Weight (WFJ2100) | 10kg |
| Weight (WFZUV2100) | 12.5kg |
The wavelength range for the WFJ2100 Visible Spectrophotometer is 325—1000 nm.
The wavelength range for the WFZUV2100 UV-Visible Spectrophotometer is 200—1000 nm.
The instrument should be placed in an environment with a room temperature of 5—35°C and relative humidity not exceeding 85%.
The power supply must be within ±10% of the rated voltage and frequency variation within ±1HZ.