The Illumina MiSeqTMDx gene sequencer user manual provides installation, operation, maintenance and safety information.
This manual describes the MiSeqTMDx gene sequencer, its hardware components, required laboratory environment, and the software workflow for DNA sequencing. It includes detailed installation steps, electrical and network requirements, and safety warnings for safe operation.
The document also covers performance specifications, routine maintenance, troubleshooting procedures, and a list of accessories. All instructions are intended for qualified personnel working with the instrument in an in‑vitro diagnostic setting. This manual is written in Chinese.
| Storage temperature range | -10 °C to 40 °C |
|---|---|
| Operating temperature range | 19 °C to 25 °C |
| Operating humidity | 30 %–75 % non‑condensing RH |
| Power input | 100–240 V AC @ 50/60 Hz |
| Power consumption | 400 W |
| Heat output | 400 W (1,364 Btu/h) |
| Maximum altitude | < 2000 m |
The instrument may be stored at temperatures between –10 °C and 40 °C.
It requires an AC input of 100–240 V at 50/60 Hz and consumes up to 400 W.
Relative humidity should be kept between 30 % and 75 % non‑condensing.