AO-07 Medical Oxygen Sensor is a medical oxygen sensor with long life, fast response, high accuracy, high reliability, etc. It is suitable for applications such as oxygen analyzer.
This document provides operating instructions, technical specifications, and safety precautions for the Aosong AO-07 medical oxygen sensor. It details installation requirements, calibration procedures, cleaning methods, and environmental limits. The sensor is designed for medical applications such as ventilators and anesthesia equipment, with features like temperature compensation and fast response. Safety warnings include avoiding MRI environments and not using the sensor for applications that may cause personal injury. This manual is written in Chinese.
| Measurement Range | 0 – 100 Vol.% O2 |
|---|---|
| Response Time (T90) | <15 s |
| Accuracy | meets ISO 80601-2-55 requirements |
| Operating Temperature Range | -20℃ - +50℃ |
| Operating Pressure Range | 0.5 Bar - 2.0 Bar |
| Weight | 42 g (nominal) |
| Expected Operating Life | 20℃ 1.5 x 106%O2 hours |
The operating temperature range is -20℃ to +50℃.
At 20℃, the expected operating life is 1.5 x 106%O2 hours.
Place it in quarantine for 24 hours and perform a follow-up check with a 2-point calibration.
No, the sensor is not suitable for sterilization by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide.