Thorlabs PDP90A 2D Lateral Effect Position Sensor User Guide

The PDP90A is a two-dimensional optical sensor that measures X/Y position. The sensor can be used with Thorlabs' KPA101 K-Cube Auto-Aligner.


Thorlabs PDP90A 2D Lateral Effect Position Sensor User Guide - cover page
Brand
THORLABS
Category
sensor
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
1 MB
Published
27 October, 2023
Updated
27 October, 2023
Model
PDP90A
MD5 checksum
4D9437ABDCC923B8762C4965101778A9

About this manual

The PDP90A from Thorlabs is a two-dimensional, tetra-lateral position-sensing detector that provides X- and Y-axis positional information. It is designed to work with the Thorlabs KPA101 K-Cube Auto-Aligner and can be used for general alignment and precise position measurements. The sensor supports spot sizes up to 9 mm, with a recommended spot between 0.2 mm and 7 mm.

Setup instructions include unpacking and mounting the detector, removing the protective lens film, connecting to the KPA101, and sending a beam onto the active area. The manual explains operation, position calculation, resolution, and specifications such as wavelength range and output voltage.

What's inside

  • Warning Symbol Definitions
  • Description
  • Specifications
  • Drawing
  • Regulatory
  • Thorlabs Worldwide Contacts

Specifications

Wavelength Range320 - 1100 nm
Peak Responsivity0.6 A/W at 960 nm
Transimpedance Gain100 kV/A
Photocurrent (Max)40 µA
Output Voltage Range-4 to +4 V (Pins 1 & 2); 0 to 4 V (Pin 3)
Bandwidth15 kHz
Spot SizeØ0.2 to Ø7 mm recommended; Ø9 mm max
Power Requirements±5 VDC ± 5%, 35 mA

Frequently asked questions

What is the position resolution of the PDP90A?

The position resolution is 0.75 µm. It is calculated from the detector resistance length and output noise, with best results at a SUM output of 4 V.

What spot size is recommended for the PDP90A?

The recommended spot size is between 0.2 mm and 7 mm, and the spot should be within the center 80% of the detector surface. The maximum spot size is 9 mm.

What is the maximum photocurrent before saturation?

The maximum photocurrent before saturation is 40 µA. The amplifier would saturate before the photodiode itself.