Sysmex CS-5100 Application guide for platelet function test by automated aggregation with epinephrine

This device can measure platelet function, and it is composed of Sysmex CS-5100 system and Epinephrine kit


Sysmex CS-5100 Application guide for platelet function test by automated aggregation with epinephrine - cover page
Brand
Sysmex
Category
Professional medical equipment
Document type
Application Note
Language
English
Pages
19
File format
PDF
File size
1.2 MB
Published
27 October, 2023
Updated
27 October, 2023
MD5 checksum
71A6FE8E0C2FB32D3A596D9B4DEECE8D

About this manual

This application guide describes the parameters and procedure for measuring epinephrine-induced platelet aggregation using the Epinephrine kit (AG002K) on the Sysmex CS-5100 system with software version 00-13. It is the valid source of information for the assay application in the Sysmex CS-5100 system.

The guide covers test principle, reagent preparation and stability, performance characteristics, result interpretation, and analyzer setup including assay group configuration, calculation methods, and formula settings for Epi% measurements.

What's inside

  • Test Principle
  • Test synopsis
  • Reagents and material required
  • Preparation and stability of reagents
  • Recommendations
  • Results
  • Performance and characteristics
  • Comparison of Methods
  • Setup of test
  • Formula setting

Specifications

Measurement range0-100% of maximum aggregation
Measurement time600 sec at 37°C
Wavelength660 nm
Sample volume140 µL
Reference range68.8 – 99.8 % of maximum aggregation

Frequently asked questions

What is the test principle for epinephrine-induced platelet aggregation?

Epinephrine binds to alpha2-adrenergic receptors on platelets, inducing an initial reversible aggregation followed by a second wave from ADP release and thromboxane A2 synthesis.

How are the results expressed?

Results are expressed in % of maximum aggregation, and duplicate measurements are recommended.

What software version is the guide based on?

The application guide is based on the Sysmex CS-5100 system with software version 00-13.

What is the acceptable coefficient of variation for normal samples?

The coefficient of variation for normal samples must be less than 20%.