Instructions for Mercury Freedom System calibration
This technical bulletin provides step-by-step calibration instructions for the Thermo Fisher Mercury Freedom System. It is marked Rev. 09/2015 and includes advanced instrument and system calibration procedures, manual daily instrument and system calibration, a system linearity check, and a system integrity check.
The procedures use MAIN MENU selections such as INSTRUMENT CONTROL, CALIBRATION, GAS MODE, SERVICE MODE, and SAMPLE. Notes describe coefficient checks, automatic calibration behavior, and a chlorine pressure requirement for the system integrity check.
During an instrument zero, the document says to wait 10 minutes.
For system zero or system span, wait at least 15 minutes until concentration readings are stable.
For manual daily instrument calibration, perform the Advanced Instrument Calibration; for manual daily system calibration, perform the Advanced System Calibration.
The system integrity check requires chlorine gas to be turned on and pressure set between 10 and 15 psi.