Barnstead LabTower EDI System Quick Reference Guide
This document serves as a quick reference guide for the Thermo Fisher LabTower EDI System, providing essential information on system components, consumable replacement schedules, and pretreatment recommendations. It includes specific details on replacement intervals for items such as RO membranes, ultrapure cartridges, and UV lamps based on usage hours and water quality.
| Storage Tank Volume | 100 L |
|---|---|
| Recirculation Pump Flow | 1.5 LPM at 1.2 - 2 bar |
| RO Membrane Reduction | up to 99 % |
| UV Lamp Life | Up to 2 years |
| Ultrapure Cartridge Life | Up to 12 months |
Replace the RO membrane when the permeate conductivity limit is reached or when the display shows 'Lim.val. permeate'.
It indicates that the purity has reached the pure water limit value, requiring the replacement of the ultrapure cartridge.
The UV lamp should be replaced every 2 years or when the UV duration is displayed, whichever comes first.
The ultrapure cartridge should be changed at a minimum of once a year.