Thermo Fisher LabTower EDI System Quick Reference Guide

Barnstead LabTower EDI System Quick Reference Guide


Thermo Fisher LabTower EDI System Quick Reference Guide - cover page
Brand
Thermo Fisher
Category
Other
Document type
Quick Start Guide
Language
English
Pages
2
File format
PDF
File size
348 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
3F552F159B8E38E5A773858526090034

About this manual

This document serves as a quick reference guide for the Thermo Fisher LabTower EDI System, providing essential information on system components, consumable replacement schedules, and pretreatment recommendations. It includes specific details on replacement intervals for items such as RO membranes, ultrapure cartridges, and UV lamps based on usage hours and water quality.

What's inside

  • LabTower EDI Consumables Replacement Information
  • Display Functionality Keys and Readout
  • Common Alerts
  • Maintenance
  • Pretreatment
  • System Components
  • Operating Modes

Specifications

Storage Tank Volume100 L
Recirculation Pump Flow1.5 LPM at 1.2 - 2 bar
RO Membrane Reductionup to 99 %
UV Lamp LifeUp to 2 years
Ultrapure Cartridge LifeUp to 12 months

Frequently asked questions

When should the RO membrane be replaced?

Replace the RO membrane when the permeate conductivity limit is reached or when the display shows 'Lim.val. permeate'.

What does the 'Lim.val.pure.w' alert indicate?

It indicates that the purity has reached the pure water limit value, requiring the replacement of the ultrapure cartridge.

How often should the UV lamp be replaced?

The UV lamp should be replaced every 2 years or when the UV duration is displayed, whichever comes first.

What is the minimum replacement frequency for the ultrapure cartridge?

The ultrapure cartridge should be changed at a minimum of once a year.