This manual provides instructions for the ROSS® Process Combination pH Electrode
This user guide provides instructions for preparing, installing, and maintaining the ROSS® Process Combination pH Electrode (model 2001SC) for online pH measurement. It includes details on electrode preparation, attaching reference filling solution, installing the electrode into the Orion flow cell (2001FC), and best practices for calibration and maintenance. The electrode is designed for stability and accuracy in varying temperatures, using a proprietary internal system to avoid drift and failure common in conventional electrodes.
The guide also covers safety precautions, such as wearing gloves and avoiding contamination, and includes steps for venting the reference solution bottle and replacing tubing assemblies every 3–6 months to ensure optimal performance.
| Model Number | 2001SC |
|---|---|
| Reference Filling Solution | ROSS reference filling solution, Cat. No. 2001FS |
| Flow Cell Compatibility | Orion flow cell, Cat. No. 2001FC |
| Electrode Slope | 92 to 102% |
The model number is 2001SC, as stated in the document.
Use only ROSS reference filling solution, Cat. No. 2001FS, as other solutions may cause failure.
Replace the tubing assembly every three to six months.