This file is an operating manual for Front Load Single and Dual Sensors PN 074-156N.
This is the INFICON Operating Manual for Front Load Single and Dual Sensors, document PN 074-156N. It introduces the single and dual sensor configurations and their parts, and covers shutter modules, feedthroughs, installation requirements, and drawings.
The manual also describes sensor installation, pre-installation sensor checks with various deposition controllers and monitors, tube bending, solenoid valve assembly installation, maintenance and spare parts. It includes crystal replacement instructions, sensor maintenance, and Crystal Fail checks using the Crystal Sensor Emulator.
| Feedthrough bolt size | 2.54 cm (1 in.) |
|---|---|
| ConFlat feedthrough size | CF 40 (2-3/4 in. ConFlat) |
| Crystal Sensor Emulator part number | 760-601-G2 |
| Crystal Sensor Emulator fundamental frequency | 5.5 MHz |
| Crystal Life display with emulator, incrementing from 0%, 1 MHz shift | approximately 45% |
The Crystal Sensor Emulator contains a quartz crystal with a fundamental frequency of 5.5 MHz.
It should read approximately 45%.
It should read approximately 60%.