Key facts
Start here

JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual
The JOFRA ETC-125 A / 400 A / 400 R user- and reference manual describes how to set up, operate, service and maintain the temperature calibrator. It also includes safety symbols, technical specifications and a list of accessories.
The manual gives thermal specifications for the ETC-125 A, including maximum temperature 125°C / 257°F, display resolution and stability, plus electrical and mechanical specifications for the ETC-125 A and ETC-400 A / R. It lists standards for EMC and low-voltage directives and accessories such as insertion tubes, carrying case and software.
JOFRA 400 A specifications
| Calibration interval | 1 year |
|---|---|
| Warm‑up time | 5 minutes |
| Stable temperature range for warm‑up | 18 °C to 28 °C |
| Minimum stabilization time if outside range | 3 hours |
| USB virtual COM port name | AMETEK Advanced Signal Calibrator |
| Calibration menu password | 900 |
Source: JOFRA ASC-400 Advanced Signal Calibrator Adjustment Manual
All JOFRA 400 A documents by type
Disassembly, diagnostics and part numbers, for repair.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What password is required to access the calibration menu?
The password is 900.
From: JOFRA ASC-400 Advanced Signal Calibrator Adjustment Manual
Which test equipment is recommended for adjusting the ASC-400?
A Fluke 8508A digital multimeter, a Fluke 5520A calibrator, and appropriate banana leads are listed as recommended equipment.
From: JOFRA ASC-400 Advanced Signal Calibrator Adjustment Manual
What is the maximum temperature of the ETC-125 A?
The maximum temperature is 125°C / 257°F.
From: JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual
What is the display resolution?
The display resolution is 1° or 0.1°C / 1.8° or 0.18°F.
From: JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual
What ambient temperature is used for the thermal specifications?
All thermal specifications are given with an ambient temperature of 23°C/73.4°F ± 3°C/5.4°F.
From: JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual
Which EMC standard is cited for the calibrators?
The manual cites EN 61326-1: 2013 for EMC requirements.
From: JOFRA ETC-125 A / 400 A / 400 R Temperature Calibrator User and Reference Manual