GE Healthcare issues safety notice: Batteries in certain Vivid ultrasound systems may fail and emit smoke or catch fire.
This document is an urgent field safety notice from GE Healthcare regarding certain legacy Vivid ultrasound systems. It warns that if the internal batteries are not replaced every two years, they may fail, potentially emitting smoke or catching fire. The notice states that there have been no reported injuries but emphasizes the need for immediate action to replace the batteries.
| Battery replacement interval | every 2 years |
|---|---|
| Minimum battery runtime | 30 minutes |
| Expected new battery runtime | 60 minutes |
The battery must be replaced if it is two years old or if it cannot power the system for more than 30 minutes.
Yes, you can continue to use your device, but you must follow the safety instructions and replace the battery.
Instructions for replacement are found in the corresponding Service Manual, which can be accessed via a provided link or by contacting GE Healthcare Service.