GE Healthcare Vivid S5/S6/i/q Battery Safety Supplement

GE Healthcare issues safety notice: Batteries in certain Vivid ultrasound systems may fail and emit smoke or catch fire.


GE Healthcare Vivid S5/S6/i/q Battery Safety Supplement - cover page
Brand
GEHealthcare
Category
medical technology
Document type
Service Manual
Language
English
Pages
8
File format
PDF
File size
264 KB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
D3EEC4E74B9F9F9548C7C684FC5ED894

About this manual

This document is an urgent field safety notice from GE Healthcare regarding certain legacy Vivid ultrasound systems. It warns that if the internal batteries are not replaced every two years, they may fail, potentially emitting smoke or catching fire. The notice states that there have been no reported injuries but emphasizes the need for immediate action to replace the batteries.

What's inside

  • Safety Issue
  • Actions to be taken by Customer/User
  • Affected Product Details
  • Acknowledgement Response
  • Battery Safety Precautions
  • Battery replacement
  • Contact Information

Specifications

Battery replacement intervalevery 2 years
Minimum battery runtime30 minutes
Expected new battery runtime60 minutes

Frequently asked questions

What are the conditions for replacing the battery?

The battery must be replaced if it is two years old or if it cannot power the system for more than 30 minutes.

Can I continue to use the device while waiting for the battery replacement?

Yes, you can continue to use your device, but you must follow the safety instructions and replace the battery.

How do I find the specific service manual for my model?

Instructions for replacement are found in the corresponding Service Manual, which can be accessed via a provided link or by contacting GE Healthcare Service.