GE Healthcare Sephacryl S-100 to S-500 High Resolution Instructions

Instructions for preparing medium suspension and packing chromatography columns using GE Healthcare Sephacryl HR gel filtration media.


GE Healthcare Sephacryl S-100 to S-500 High Resolution Instructions - cover page
Brand
GEHealthcare
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
217 KB
Published
18 October, 2026
Updated
18 October, 2026
MD5 checksum
C509E4B8D5366AF72E8FC24E8A4F43F3

About this manual

This document provides detailed instructions for the preparation of Sephacryl™ HR gel filtration media suspensions and the subsequent packing of chromatography columns. It outlines a specific two-step packing method designed to ensure the medium is most tightly packed at the top of the column, thereby improving resolution compared to traditional methods.

What's inside

  • Introduction
  • To prepare the medium suspension
  • To pack the column
  • Recommended packing flow rates
  • Efficiency test (plate number)
  • Medium characteristics

Frequently asked questions

What is the recommended procedure to improve resolution when packing the column?

You should pack the column in two steps, then insert a second adaptor at the top, and finally turn the column upside-down so the sample is applied in the most tightly packed zone.

How much medium slurry should be measured out before diluting it with eluent buffer?

You must measure out a volume of medium slurry that is 1.5 times the desired packed medium volume.

What are the recommended packing flow rates for an XK 26/70 column using aqueous buffers at room temperature?

The recommended flow rate is 150 ml/h for the first step and 300 ml/h for the second step.