Instructions for preparing medium suspension and packing chromatography columns using GE Healthcare Sephacryl HR gel filtration media.
This document provides detailed instructions for the preparation of Sephacryl™ HR gel filtration media suspensions and the subsequent packing of chromatography columns. It outlines a specific two-step packing method designed to ensure the medium is most tightly packed at the top of the column, thereby improving resolution compared to traditional methods.
You should pack the column in two steps, then insert a second adaptor at the top, and finally turn the column upside-down so the sample is applied in the most tightly packed zone.
You must measure out a volume of medium slurry that is 1.5 times the desired packed medium volume.
The recommended flow rate is 150 ml/h for the first step and 300 ml/h for the second step.