GE Healthcare issued an urgent field safety notice to remind users that CT, PET, or NM tables may drop during specific servicing activities, which may cause injury to service personnel.
GE Healthcare issued an urgent field safety notice regarding potential table drop during specific servicing activities on affected CT, PET, and NM tables. The issue stems from unclear guidance in service manuals, not a defect in the tables themselves. Users are advised to contact GE Healthcare Service before performing any table servicing to receive updated instructions. The updated service manuals are expected by the end of June 2022 and can be accessed via the provided link.
The tables may drop during specific servicing activities if the wrong screws are removed without appropriate table supports in place, potentially causing injury to service personnel.
Contact GE Healthcare Service prior to any table servicing activities to receive guidance on safe servicing procedures.
The updated service manuals are planned to be released by the end of June, 2022.