This is a PDF file about GE Healthcare equipment installation requirements
This document outlines installation requirements for the GE Healthcare B7864PZ / P5064PS / E4502F Three Phase, 14.4 kVA UPS. It includes site preparation, power, and facility modifications needed for safe delivery and operation. The UPS provides backup power to maintain critical scanner functions during outages and protects electronics from voltage anomalies. Installation must comply with GE Healthcare’s specifications and local regulations.
The UPS features true double-conversion technology, automatic voltage/frequency selection, and a graphical display for status monitoring. It is tested and approved for use with most CT & PET/CT scanners, with compatibility verified by the factory. Mobile applications require additional mounting kit B7864MK.
| Rating | 14.4 kVA |
|---|---|
| Operating Temperature | 10-40°C |
| Input Voltage Range | Three-phase; 102-132V / ph |
| Input Frequency Range | 45-65 Hz |
| Output Frequency | 50 or 60 Hz, auto-sensing |
| Voltage Regulation | <3% steady-state for all conditions of line and load |
| Battery Backup Time | >10 minutes typ. |
| Weight | 620 lbs (277 kg) |
| Efficiency | >90% typ. |
Battery backup time is greater than 10 minutes typical.
Yes, but mobile applications require the addition of mobile kit B7864MK for secure mounting.
It meets IEC 62040-1-1, UL1778, cUL, CAN/CSA C22.02 NO.107.3 safety standards.