GE Healthcare issues urgent field safety notice to remind users to update the service procedure of Proteus XR/a and Revolution XR/d systems for collimator re-installation errors.
This document is an urgent Field Safety Notice from GE Healthcare concerning the Proteus XR/a and Revolution XR/d X-ray systems. It addresses potential errors in collimator re-installation procedures that could lead to equipment failure and patient or operator injury. The notice states that existing service manuals at some sites may not contain the latest instructions required to ensure proper re-installation.
Servicing instructions may be outdated, leading to collimator re-installation errors that can cause the collimator to fall and injure a patient or operator.
The Preventive Maintenance Manual must be version 2273022-100 Rev 30 or greater (PM 012), and the Installation Manual must be version 2273022-100 Rev 30 or greater (IST 014, IST 015).
The Preventive Maintenance Manual must be version 2296990-100 Rev 9 or greater (Section 2-5-2), and the Installation Manual must be version 2357602-100 Rev 10 or greater (Section 7-7, Section 7-9).