GE Healthcare has updated the Service Manuals for the products listed below to update an incorrect torque value.
GE Healthcare issued an urgent safety notice to update torque values in service manuals for specific CT and PET scanners. The error could cause over-torquing during power supply replacement, risking bolt failure and serious injury. Users must ensure service providers use the latest manual. Affected models include LightSpeed Ultra, LightSpeed Plus, LightSpeed QX/i, LightSpeed 16, Discovery ST, and Discovery LS. The updated manuals are available online via GE Healthcare’s documentation library.
The service procedure for replacing a power supply specifies incorrect torque values, which could lead to over-torquing and potential bolt failure during operation.
The latest versions are available online at http://apps.gehealthcare.com/servlet/ClientServlet?REQ=Enter+Documentation+Library.
No further action is needed on your power supply if serviced by GE Healthcare, but you must update your service manual.