GasAlertMicro 5/PID/IR Quick Reference Guide introduces the GasAlertMicro 5/PID/IR multi-gas detector. It includes: safety information before use, components, display elements, connecting gas cylinders, performing a general test, calibration, operator option menu, maintenance, replacing batteries, replacing sensors or sensor filters, replacing pump filters and so on
This guide provides essential information for operating the GasAlertMicro 5/PID/IR multi-gas detector. It covers safety precautions, sensor handling, calibration, battery replacement, and maintenance. The detector is designed to alert users to hazardous gas levels above user-defined thresholds. It includes safety warnings, calibration procedures, and instructions for replacing sensors and batteries. The guide also includes warranty and liability information, as well as contact details for BW Technologies by Honeywell.
The detector supports multiple languages including English, French, German, Portuguese, and Spanish. It is intended for use in potentially explosive atmospheres with oxygen levels not exceeding 20.9% (v/v). Regular calibration is required, at least every 180 days, to ensure accurate readings. The guide advises users to follow all safety instructions and to consult the full user manual for complete operating instructions. This manual is written in Portuguese.
| Warranty Period | 2 years |
|---|---|
| Calibration Frequency | Every 180 days (6 months) |
| Oxygen Level for Calibration | 20.9% (v/v) |
| Fuel Sensor Calibration | 50% LEL of methane |
The product is guaranteed to be free of material and manufacturing defects for two years from the date of shipment to the buyer.
Sensors should be calibrated regularly, at least once every 180 days (6 months).
Calibrate only in a safe area free of hazardous gases with an atmosphere of 20.9% oxygen.
The fuel sensor is factory-calibrated to 50% LEL of methane.