G1010 MkII MTL Oxygen Analysers Instruction Manual
This manual provides instructions for installing, commissioning, and maintaining the MTL G1010 MkII oxygen analyser. It includes specifications for display, output ranges, cell life, sample flow, and environmental conditions. The device is microprocessor-controlled and features user-programmable alarms and analogue outputs. It supports various sensor types for measuring oxygen concentrations from 100% down to a few ppm.
Installation requires proper mounting, power supply, and sample connections using captive seal fittings. Commissioning involves applying power, programming settings, and applying the sample. Maintenance includes calibration and cell replacement procedures.
| Display | Multi-digit LCD - character height 12.7mm |
|---|---|
| Display and Output Ranges (G1010 and G1010R) | E cell: 0.1% to 100.0%; N cell: 100ppm to 50%; H cell: 10ppm to 10%; L cell: 1ppm to 10% |
| Stability | Better than 2% of reading per month, or 1ppm whichever is greater |
| Sample Flow | Between 100 and 300 ml/min for optimum operation. Max. 1 l/min |
| Sample Temperature | -5 to +40°C (non condensing) |
| Analogue Output | 0 to 5 volts (min. load 10k ohms) or 4 to 20mA (max. load 300 ohms) |
| Mounting | Panel-mounting with two clamps. Enclosure material: Glass-fibre reinforced Noryl. Protection level: IP40, IP54 (optional door), or IP30 otherwise. |
E cell: 0.1% to 100.0%; N cell: 100ppm to 50%; H cell: 10ppm to 10%; L cell: 1ppm to 10%.
Between 100 and 300 ml/min for optimum operation, with a maximum of 1 l/min.