This is a manual for PCB Piezotronics 699B06 Calibration shaker. It supports lifetime warranty and limited warranty. It needs routine calibration and battery replacement. If the device is damaged, you can contact customer service for repair.
The PCB Piezotronics Model 699B06 Portable Shaker Table is designed for field test applications to calibrate sensors and associated instrumentation. The device is intended for use by qualified personnel in an industrial environment and requires operation on a stable, flat surface. It features a mounting platform capable of supporting loads up to 800 grams directly, with recommendations for external suspension systems for heavier loads to reduce distortion. The unit operates using a supplied 4 GB flash drive for loading CALROUTE files, and users are advised to use hearing protection during extended operation due to potential noise hazards.
| Maximum Direct Mount Load | 800 grams (28.3 ounces) |
|---|---|
| Warranty Period | TWO YEARS |
| Recommended Flash Drive Size | 4 GB |
We recommend calibrating on an annual basis, after exposure to any extreme environmental influence, or prior to any critical test.
Utilizing flash drives with more memory could cause the 699B06 to time out and restart when loading a CALROUTE.
Yes, hearing protection is recommended when operating the 699B06 for an extended amount of time.