Installation and operating guide for the PCB Piezotronics 352B70 ICP® accelerometer, including specifications, mounting, calibration and warranty information.
This manual provides step‑by‑step installation instructions, wiring details, and operating procedures for the PCB Piezotronics 352B70 ICP® accelerometer. It also covers recommended calibration practices, environmental limits, and safety precautions.
Additional sections describe warranty coverage, return‑material‑authorization (RMA) procedures, technical specifications, mounting torque values, and optional accessories that can be used with the sensor.
| Sensitivity | 1 mV/g |
|---|---|
| Measurement Range | ± 5000 g pk |
| Frequency Range (±5 %) | 0.7 to 9000 Hz |
| Operating Temperature Range | -65 to +250 °F |
| Weight | 0.15 oz (4.3 g) |
Contact your local distributor or sales representative to obtain an RMA number, mark it on the package, and include a detailed description of the problem with the return shipment.
Calibration is typically performed annually, or after exposure to severe temperature extremes, shock, load, or other environmental influences.
The recommended mounting torque for the 10‑32 thread version is 4‑5 in‑lb (45‑55 N·cm).