TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual

TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual


TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual - cover page
Brand
LaserLinc
Document type
User Manual
Language
English
Pages
22
File format
PDF
File size
685 KB
Published
20 October, 2023
Updated
20 October, 2023
MD5 checksum
F055FA67689AB2ACD6312150F6E79603

About this manual

This manual provides step-by-step instructions for operating and maintaining LaserLinc’s laser scan micrometers, including models TLAser122, TLAser222, TLAser122s, Triton312, Triton330, and TLAser400. It covers mounting, cabling, air purge use, cleaning procedures, and troubleshooting common issues. Safety precautions for laser exposure are included, along with technical specifications for each model. The manual also details preventive maintenance and how to resolve measurement errors such as 'Too Few' or 'Too Many' readings.

What's inside

  • Laser Micrometers – Theory of Operation
  • Laser Micrometers – Safety Considerations
  • TLAser122™ Laser Micrometer
  • TLAser222™ Laser Micrometer
  • TLAser122S Separated Head Laser Micrometer™
  • Triton312/330 Triple Axis Laser Micrometer™
  • Preventive Maintenance & Troubleshooting
  • TLAserX22/3xx Cleaning and Routine Maintenance
  • Troubleshooting and FAQs
  • TLAser400™ PCI Micrometer Interface Card

Specifications

Measurement range.003” - .85” (.075mm – 22mm)
Resolution.000001” (.025µm)
Two-second measurement repeatability± .000005” (± .125µm)
Single scan measurement repeatability± .0001” (± 2.5µm)
Accuracy.00002” ± .01% of the maximum measurement size
Scan Rate400 scans / second
Ambient Temp.45F-110F
Power requirements from PC+12VDC 125mA, +5VDC 100mA, -5VDC 75mA
Laser ClassClass II
Laser wavelength675nm
Weight2.6 lbs. (1.2kg)
Dimensions10”L x 1.25”W x 4”H

Frequently asked questions

What should I do if Total Vu reports 'Too Few: 4'?

Check that the part is in the field and illuminated by the laser. Check each axis via the test port for correct waveform.

What if Total Vu reports 'No Scan'?

Check that the scanner is powered and the data cable is connected to both the scanner and the TLAser400.

Download PDF (685 KB)

More LaserLinc Manuals


View all LaserLinc manuals →