User Manual 44 pages · 1.4 MB · EN

This manual begins with safety and electromagnetic compatibility information, outlining precautions such as avoiding explosive atmospheres, keeping all covers installed, and using only the specified accessories and cables. It also describes the theory of operation, intended use, and compliance details for the laser scan micrometers.

Later sections cover hardware installation, mounting, connections, preventive maintenance, troubleshooting, and step‑by‑step instructions for installing and configuring the Total Vu software that accompanies the devices.

Read online

LaserLinc TLAser400 specifications

Measurement range.003” - .85” (.075mm – 22mm)
Resolution.000001” (.025µm)
Two-second measurement repeatability± .000005” (± .125µm)
Single scan measurement repeatability± .0001” (± 2.5µm)
Accuracy.00002” ± .01% of the maximum measurement size
Scan Rate400 scans / second
Ambient Temp.45F-110F
Power requirements from PC+12VDC 125mA, +5VDC 100mA, -5VDC 75mA
Laser ClassClass II
Laser wavelength675nm
Weight2.6 lbs. (1.2kg)
Dimensions10”L x 1.25”W x 4”H

Source: TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual

All LaserLinc TLAser400 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What should I do if Total Vu reports 'Too Few: 4'?

Check that the part is in the field and illuminated by the laser. Check each axis via the test port for correct waveform.

From: TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual

What if Total Vu reports 'No Scan'?

Check that the scanner is powered and the data cable is connected to both the scanner and the TLAser400.

From: TLAserX22/3xx, TLAser400 Laser Scan Micrometers & interface card (122/222/122s, 312, 330, 400) Operator’s Manual

What safety precautions must be observed when operating the laser scan micrometer?

Do not operate the scanner in an explosive atmosphere or where flammable gases or fumes may be present, keep all covers installed during operation, and use only the accessories and cables specified in the installation instructions.

From: LaserLinc TLAserX22/Triton312/330/TLAser400 Operator’s Manual